■ Product Overview
The analysis of trace components in liquid hydrocarbon is always a technical problem in production. The more feasible method is to gasification with cylinder connected with capillary constant temperature water bath and to analyze by decrement method. Introduced the "flash" and "and so on gasification" and so on the new concept, adopt computer control, assisted with high-precision flow sensor precision measuring gas flow rate and cumulative volume, successfully solved such as olefins in the trace composition analysis of the technical problems in production, and with its simple operation, accurate measurement of the personage inside course of study.
■ Functional features
▲ high-definition 7-inch color touch screen, full-screen touch buttons, Chinese/English interface.
▲ 32-bit high-performance microprocessor, embedded Linux operating system.
▲ Temperature control using PID self-tuning technology, accurate temperature control.
▲ High precision inlet flow sensor, sample vaporized flow rate control is accurate.
■ Technical parameters
▲ Temperature measurement: platinum resistance
▲ Metal bath temperature: not less than 60℃
▲ Temperature control accuracy: ±0.1℃
▲ Temperature resolution: 0.01℃
▲ Flow rate measurement: flow sensor
▲ Flow rate range: 5.0L/M (adjustable)
▲ Flow rate control accuracy: 1%F.S
▲ Flow rate resolution: 0.01L/M screen
▲ Volume measurement: flow rate * time, calculated
▲ Volume resolution: 0.01L
▲ Quality measurement: volume * density, calculated
▲ Mass resolution: 0.001g
▲ Display: 800*480 high-definition color LCD touch
▲ Power supply: AC 220V±22V
▲ Power supply: 400VA
▲ Frequency: 50Hz±2.5Hz
▲ Ambient temperature: 10℃~35℃
▲ Ambient humidity: ≤85%RH
▲ Weight: 6kg
▲ Size: 230mm×400mm×325mm