Notes for kinematic viscosity tester test

AddTime:2021-09-10 09:20:42   Views:     【 Big Mid Small 】   Print   Close

1, when using the instrument, the instrument shell must be well grounded;

2. Heating control switch shall not be opened before the constant temperature tank is loaded with transformer oil;

3. The constant temperature medium in the bathtub shall not be lower than half of the upper part of the expanding part of the viscometer;

4, continuous use should pay attention to whether the transformer oil level in the constant temperature tank is reduced, if it is too low should be added at any time, if there are impurities in the transformer oil should be processed in time to exclude or replace the transformer oil;

5. It is forbidden to unplug or plug each connection line without unplugging the power plug;

6. The constant of capillary viscometer should be backed up safely;

7. Pay attention to replenish cleaning liquid and discharge waste oil in waste liquid storage, otherwise serious failure or serious accident will occur to the instrument;

8. Take 15mL ~ 17mL samples from the beaker and pour them into the test tube from the mouth of the cup. If too much of the test liquid will seal the lower port of the capillary, affecting the work of the viscometer. The sample can not be loaded too little, otherwise it will lead to the formation of vacuum measurement;

9, try liquid pump speed by vacuum pressure regulating valve adjustment, adjustment method is to loosen the adjusting screw counterclockwise, then in the process of smoking on the screw clockwise slowly turn into, until the pumping speed on the right, not before testing the screw, otherwise the pumping speed too fast, may be outside the tube of viscosity for test solution extraction, general modulated before delivery;

10, the instrument can use two solvents to clean the viscosity tube, in principle, cleaning 1 should choose to test samples of good solubility solvent (petroleum ether, etc.), cleaning 2 should choose to make the viscosity tube easy to dry solvent (anhydrous ethanol, etc.). Generally speaking, when doing oil testing, the cleaning liquid 1 tube is connected, and a cleaning liquid can be used; Do polymers and plastics and other solution reagents, must be the cleaning liquid 1 and cleaning liquid 2 selection and connection;

Note: Cleaning liquid 1 and 2 are in independent bottles, placed outside the back of the instrument host.

Note: this instrument uses a cleaning solution to measure the kinematic viscosity of oil products, generally 30 ~ 60 petroleum ether or 120 solvent oil.

11. In daily use of the instrument, it is necessary to observe whether there is more waste liquid in the cleaning liquid bottle, and pour it frequently. Do not make the bottle full, which will affect the normal operation of the vacuum system; After pouring the waste cleaning liquid, be sure to put the cap on the bottle, can not leak;

12, do not press the LCD; Do not use sharp objects to scratch, scratch the display screen;

13. It is normal that the color of the display screen sometimes changes with the ambient temperature and Angle of view;

14. After the experiment, turn off the power switch and unplug the power plug;

15, after the use of the instrument, pull out the cleaning liquid catheter, remove the cleaning liquid bottle, and then empty the instrument again for reuse;